4-8 FT-100/CI and FT-170/CI Operation Guide
Document 40275-0913
4.6 Verifying That the FT-170/CI X-Ray
On/Off Switch Works Correctly
4.6.1 Steps
Only the FT-170/CI includes the X-ray on/off switch, which gives
you manual control of the X-ray inspection beam.
1. Ensure that the conveyor running.
2. Turn off the X-ray on/off switch.
3. Does the red safety light turn off?
If yes, go to the next step, as the X-ray on/off switch is
working correctly.
If no, discontinue using the FT-170/CI and immediately
contact your service department, as the X-ray on/off switch
is not working correctly.
4. Turn on the X-ray on/off switch.
5. Does the red safety light turn on?
If yes, stop, as the X-ray on/off switch is working correctly.
If no, turn off the X-ray on/off switch, replace the red
safety light bulb, and go to the next step.
6. Turn on the X-ray on/off switch.
7. Does the red safety light turn on?
If yes, stop, as the X-ray on/off switch is working correctly.
If no, discontinue using the FT-170/CI and immediately
contact your service department, as the X-ray on/off switch
is not working correctly.