DMP10 / DMP20 Settings for measurement
To change the acquisition method for the air reference value
1. > Batch Modify >
OK
> More Settings >
OK
> Air Reference >
OK
> set service
code (159) > OK > set desired acquisition method >
OK
Parameter is activated:
2. Exit the menu: , the setting has now been made.
7.6 Assigning a different calibration to the opened batch
To assign a different calibration to the batch
> Batch Modify > OK > More Settings > OK > Calibration Assign > OK > select
the desired calibration from the list > OK
A different calibration has now been assigned.
[ ]
[ ]