79
DMP10 / DMP20 Service
14 Service
14.1 Probe – count rate recording/display
For error analysis, it can be helpful to view the count rates delivered directly from
the probe. Our experts can use the count rates shown in the display to draw conclu-
sions about the status of the probe, for example, and propose potential solutions.
The count rates are measured and displayed in a service area, and this does not af-
fect the batch.
You can also export the count rates, although only during the measurement (online),
to an Excel® table via USB and Tactile Suite.
Before you start
The probe with count rates to be checked is connected to the gauge.
For a data transfer to Excel®, the software programs Tactile Suite and MS Excel
are installed on the PC and opened. The USB cable from the scope of supply is
ready.
For the data transfer, use the USB cable from the scope of supply or any other
commercially available USB cable with a C/A-type plug; maximum usable cable
length: 3 m (118.1 ")
The link for downloading Tactile Suite and additional information can be
found on page 67.
To display/export the count rates
1. > Gauge Settings > OK > Probe - Menu > OK > Probe Raw Values > OK > Coun-
trate - Check > OK
2. Carry out one of the following measuring methods:
Individual measurements
Continuous count rate recording:
Place the gauge on the surface and press RUN; the gauge displays the
continuously recorded count rates.
[ ]