33
DMP10 / DMP20 Settings for measurement
7 Settings for measurement
In order to measure, you need a batch (measuring file). Settings, such as those for
the measurement procedure are saved in the batch, e.g. specification limit monitor-
ing and measuring mode. The link to the calibration to which the measurements are
referenced is also stored in the batch file.
A description of the possible batch settings you have to make befo re a mea-
surement can be found below.
A description of the batch settings, which you can also make after a mea-
surement, can be found from page 69.
7.1 Activating/deactivating specification limit monitoring
To activate specification limit monitoring and set the specification
limits
1. Select the specification limits to be monitored: > Batch Modify > OK > Limits
> OK > select desired specification limit parameter > OK.
Parameter is activated:
The selected specification limit(s) with the currently set values appear in the
specification limit menu.
2. Set the specification limit: select specification limit > OK > set desired value >
OK.
Repeat step 2 for the second specification limit, provided you have selected
Both in step 1.
3. Exit the menu: , the settings are now made
In the Measurement view, lines with the entered specification limits appear
above and below the measured reading or a line appears below the measured
reading with the entered limit value.
[ ]