38
Measure DMP10 / DMP20
8 Measure
8.1 Notes on measuring
The information for precision (remaining deviations from precision to mea-
sured reading) and repeatability (repeated standard deviation) applies for the
ambient and reference piece temperatures at the time of the calibration. The
values for precision and repeatability may be higher than the values indicated
in the data sheet if the temperatures during the measurement deviate from the
ambient and reference piece temperatures during the calibration.
As a basic rule: If the specified precision is not achieved during the measure-
ment, you must check the calibration recalibrate the probe if necessary.
If all signal functions in the gauge are activated, the gauge reports the following
states:
Beeps
Short beep = measured reading acquisition complete, second beep in addi-
tion if the specification limit is violated
Light signals
Specification limit monitoring not active:
Green = measured reading acquisition successful
Specification limit monitoring active:
Red = measured reading outside the specification limits, the measured
reading itself is also displayed in red.
For the following measurement situation, we recommend setting the acquisi-
tion of the air reference value to Static:
Measurements in cavities and pipes if the minimum lift off distance of
4 x the probe measurement range cannot be respected, 43