18 Settings
18.2 Measurement settings FLUXUS F501
2020-06-30, UMFLUXUS_F501V1-3EN
124
An upper limit for the flow velocity can be entered, see section 16.3.
• Enter zero to deactivate the flow velocity check.
• Press ENTER.
A lower limit for the flow velocity can be entered.
• Select sign to define a cut-off flow depending on the flow direction. One limit is set for the positive and negative flow velocity.
• Select absolut to define a cut-off flow independent of the flow direction. A limit is set for the absolute value of the flow
velocity.
• Press ENTER.
• Select factory to use the default limit of 2.5 cm/s (0.025 m/s) for the cut-off flow.
• Select user to enter the cut-off flow.
• Press ENTER.
If Cut-off Flow\sign and user are selected, 2 values have to be entered:
• Enter the cut-off flow.
• Press ENTER.
All positive values of the flow velocity smaller than this limit are set to zero.
• Enter the cut-off flow.
• Press ENTER.
All negative values of the flow velocity greater than this limit will be set to zero.
If Cut-off Flow\absolut and user are selected, only one value have to be entered:
• Enter the cut-off flow.
• Press ENTER.
The absolute value of all flow velocity values lower than this limit will be set to zero.
• Select the overflow behavior of the totalizers, see section 16.1.
• Press ENTER.
• Select on to keep the previous totalizer values after a restart of the measurement.
• Select off to reset the totalizers to zero after a restart of the measurement.
• Press ENTER.
Special Funct.\...\Measuring\Velocity limit
Special Funct.\...\Measuring\Cut-off Flow
Special Funct.\...\Measuring\Cut-off Flow\+Cut-off Flow
Special Funct.\...\Measuring\Cut-off Flow\-Cut-off Flow
Special Funct.\...\Measuring\Cut-off Flow
Special Funct.\...\Measuring\Quant. wrapping
Special Funct.\...\Measuring\Quantity recall