EasyManuals Logo

Flexim FLUXUS F501 User Manual

Default Icon
146 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #95 background imageLoading...
Page #95 background image
14 Data logger
FLUXUS F501 14.3 Configuration of the data logger
95
UMFLUXUS_F501V1-3EN, 2020-06-30
14.3 Configuration of the data logger
Select the menu item Special Funct.\SYSTEM settings\Storing.
Press ENTER.
Starting time
It is possible to set a starting time if it is necessary to synchronize the storing of measured values for several transmitters.
Select the moment at which the storing has to start.
Ringbuffer
The setting of the ringbuffer influences the storing of measured values as soon as the data logger is full:
If the ringbuffer is activated, the available data logger will be halved. The oldest measured values will be overwritten.
Only the data logger memory that was free during the activation will be used by the ringbuffer. If more data logger memory
is necessary, measured values in the data logger should be previously deleted.
If the ringbuffer is deactivated, the storing of measured values will be stopped.
Select ON to activate the ringbuffer.
Press ENTER.
Storage mode
Select the storage mode.
Press ENTER.
If sample is selected, the current measured value will be used for the storing and online transmission of data.
If average is selected, the average of all values measured during a storage interval will be used for the storing and online
transmission of data.
Special Funct.\SYSTEM settings\Storing
Special Funct.\SYSTEM settings\Storing\Start logger
display description
Promptly The storing starts immediately.
On full 5 min. The storing starts in the next full 5 minutes.
On full 10 min. The storing starts in the next full 10 minutes.
On quarter hour The storing starts in the next full 15 minutes.
On half hour The storing starts in the next full 30 minutes.
On full hour The storing starts in the next full 60 minutes.
Example
actual time: 09:06 am
setting: On full 10 min.
The storing starts at 09:10 am.
Special Funct.\SYSTEM settings\Storing\Ringbuffer
Special Funct.\SYSTEM settings\Storing\Storage mode
Notice!
The storage mode does not affect the outputs.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim FLUXUS F501 and is the answer not in the manual?

Flexim FLUXUS F501 Specifications

General IconGeneral
BrandFlexim
ModelFLUXUS F501
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals