11 Maintenance and cleaning
FLUXUS F501 11.2 Cleaning
89
UMFLUXUS_F501V1-3EN, 2020-06-30
11.2 Cleaning
Stainless steel housing
Clean the housing using a soft cloth and care and cleaning spray for stainless steel.
Aluminum housing
Clean the housing with a soft cloth. Do not use detergents.
Transducers
Remove traces of coupling compound from the transducers with a soft paper towel.
11.3 Calibration
If installed as recommended in an appropriate location, used cautiously and serviced conscientiously, no troubles should
appear.
The transmitter has been calibrated at factory and, usually, a re-calibration of the transmitter is not necessary.
A re-calibration is recommended if:
the contact surface of the transducers show visible wear or
the transducers were used for a prolonged period at high temperatures (several months > 130 °C for normal transducers
or > 200 °C for high temperature transducers)
In order to realize a recalibration under reference conditions either the transmitter, the transducers or the transmitter with
transducers have to be sent to FLEXIM depending on which part needs to be calibrated.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim FLUXUS F501 and is the answer not in the manual?

Flexim FLUXUS F501 Specifications

General IconGeneral
BrandFlexim
ModelFLUXUS F501
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals