6 Installation
FLUXUS G809 6.2 Transducers
47
UMFLUXUS_G809V2-2EN, 2020-06-25
Repeat the steps to fix the second rail, see Fig. 6.32.
Step 5: Installation in the transducer mounting fixture Variofix L
Press the transducers firmly into the transducer mounting fixture in the covers until the transducers are tightly fixed. The
transducer cables show in opposite directions, see Fig. 6.33.
Adjust the transducer distance displayed by the transmitter, see Fig. 6.34.
Fix the transducer cables with the strain relief clamp to protect them from mechanical strain, see Fig. 6.34.
Stick coupling foil (or some coupling compound for a short-term installation) on the contact surface of the transducers.
The coupling foil can be fixed to the contact surface with some coupling compound.
Put the covers with the transducers on the rails.
Correct the transducer distance, if necessary.
Fig. 6.31: Ratchet clasp with tension strap
1 sense of rotation
2 edge
3 lever
4 clasp screw with slot
Fig. 6.32: Pipe with 2 rails
Notice!
The arrows on the transducers and the covers have to point in the same direction.
Fig. 6.33: Installation of the transducers in the covers
1 cover
2 transducer clamping fixture
1
2