9 Measurement
9.3 Measurement settings FLUXUS H721
2022-05-15, UMFLUXUS_H721V1-5EN
110
9.3.4 Dynamic damping
If dynamic damping is activated, volatile changes in the measured values of the selected physical quantity are transmitted
through the transmitter without any time lag.
•Select On to activate the dynamic damping for the selected measuring channel.
• Press ENTER.
This display will only be indicated if dynamic damping has been activated in the menu item Special functions\
Measurement\Measurement settings\Dynamic damping.
• Enter the value for the dynamic threshold for the particular measuring channel. If zero is entered, dynamic damping will
be deactivated.
• Press ENTER.
• Enter the damping factor for the temporary damping.
• Press ENTER.
9.3.5 Input of the error delay
The error delay is the time interval after which the error value is transmitted to the output.
This display will only be indicated if the list item Edit is selected in the menu item Special functions\Dialogs/
Menus\Error delay.
If no error delay is entered, the damping factor will be used.
• Enter a value for the error delay.
• Press ENTER.
• Press and hold to return to the main menu.
The dynamic damping will only have impact on the selected physical quantity. All other physical quantities are not
dynamically damped.
Options\...\Dynamic damping
Options\...\Dynamic damping\Dynamic threshold
Options\...\Dynamic damping\Transient damping
If another physical quantity is selected, the dynamic damping has to be entered again.
Options\...\Error delay