19  Advanced functions
19.6  Remote functions FLUXUS H721
2022-05-15, UMFLUXUS_H721V1-5EN
184
Reset of measured values
• Select the list item Reset meas. val..
• Press ENTER. 
The measured value output simulates a reposing application for the duration of the signal. The actual measured flow 
velocity is ignored and the measured value is set to zero. All values of the physical quantity derived from the flow velocity 
and the values from the calculation channels also yield zero.
The transmitter continues the measurement if the condition of the remote function is no longer met.
• Select the input which is to be used to trigger the selected remote function.
• Press ENTER.
•Select No linkage in order to deactivate the remote function.
• Press ENTER.
Reset of totalizers
• Select the list item Reset totalizers. 
• Press ENTER. 
The totalizers are set to zero. The totalizers are deactivated for the duration of the signal.
Totalizing starts at zero again, as soon as the condition for the remote function is no longer met.
When the totalizers are reset to zero with the remote function, the character H is displayed next to the measured value 
during the measurement.
• Select the input which is to be used to trigger the selected remote function.
• Press ENTER.
•Select No linkage in order to deactivate the remote function.
• Press ENTER.
Stop of totalizers
• Select the list item Stop totalizers. 
• Press ENTER.
The totalizers are stopped for the duration of the signal.
Totalizing will be continued with the last registered totalizer value if the condition of the remote function is no longer met.
• Select the input which is to be used to trigger the selected remote function.
• Press ENTER.
•Select No linkage in order to deactivate the remote function.
• Press ENTER.
Activation of the FastFood mode
• Select the list item Activate FastF.
• Press ENTER.
The FastFood mode is activated for the duration of the signal. It is deactivated as soon as the condition for the remote 
function is no longer met.
This list item is only available for measuring channels and only appears if the FastFood mode was activated in menu item 
Special functions\Measurement\Measurement modes\FastFood.
• Select the input which is to be used to trigger the selected remote function.
• Press ENTER.
•Select No linkage in order to deactivate the remote function.
• Press ENTER.