19  Advanced functions
FLUXUS H721  19.7  Event triggers
189
UMFLUXUS_H721V1-5EN, 2022-05-15
19.7.3 Event trigger during the measurement
An event trigger with the switching condition MAX (x>limit), MIN (x<limit), Within range or Out of range is 
updated max. once per second to avoid a constant switching of the event trigger (in case the measured values fluctuate 
around the value of the switching condition).
An event trigger with switching condition ERR (x=fail) is activated during a measurement failure.
An event trigger of the type Non-hold is activated when the switching condition is met. It is deactivated when the 
switching condition is no longer met. But it remains activated for at least 1 s even when the switching condition is met for a 
shorter period of time.
An event trigger of the type Hold is activated when the switching condition is met. It remains activated even when the 
switching condition is no longer met.
An event trigger of the type Hold for a while is activated when the switching condition is met. The time after which 
the deactivation takes place is defined in the menu item Hold interval.
19.7.4 Status display of the event triggers
The state of the event triggers is displayed during the measurement.
• Press   to scroll through the second line from below until the event trigger status is displayed.
The status display of the event triggers is structured as follows.
Rx =   with X being the number of the event trigger and   a pictogram according to Tab. 19.6.
There is no visual or acoustic indication of event trigger switching.
Tab. 19.6: Pictograms for the status display of the event triggers
no. Function
(switching condition)
Type
(holding behavior)
current state
R=
1 MAX (x>limit) Non-hold deactivated
(false state)
2 MIN (x<limit) Hold activated
(true state)
3 Within range Hold for a while
4 Out of range
ERR (x=fail)
R1 =