7.1.3.3 Transducer cable with stainless steel conduit and stripped ends
Remove the blind plug for the connection of the transducer cable.
Insert the transducer cable into the junction box.
Fix the transducer cable by tightening the cable gland.
Connect the transducer cable to the terminals of the junction box.
Fig. 7.7: Connection of the transducer cable with plastic jacket and stripped ends
1 cap nut
2 compression part
3 basic part
4 external shield, brushed back
Tab. 7.7: Terminal assignment
terminal connection
V transducer (brown cable, marked white)
VS transducer (red cable)
RS transducer (red cable)
R transducer (brown cable)