7 Connection
7.5 Temperature probe FLUXUS H721
2022-05-15, UMFLUXUS_H721V1-5EN
88
7.5.1 Direct connection of the temperature probe
Remove the blind plug for the connection of the temperature probe.
Open the cable gland of the temperature probe. The compression part remains in the cap nut.
Push the cable of the temperature probe through the cap nut, the compression part, the basic part and the reducer.
Prepare the cable.
Shorten the external shield and brush it back over the compression part.
Insert the cable into the housing.
Screw the sealing ring side of the reducer into the transmitter housing.
Screw the basic part into the reducer.
Fix the cable gland by screwing the cap nut onto the basic part.
Connect the temperature probe to the terminals of the transmitter.
Fig. 7.19: Preparation of the temperature probe
1 cap nut
2 compression part
3 basic part
4 reducer
5 external shield, brushed back
6 sealing ring side
Fig. 7.20: Temperature probe
1–red
2 red/blue
3 white/blue
4–white
Tab. 7.16: Terminal assignment (transmitter)
terminal temperature probe
T1a...T4a red
T1A...T4A red/blue
T1b...T4b white/blue
T1B...T4B white
110 mm
7 mm
12 34
6
5
12 34