29
Table7.2Current(TRMS)
Function Range Accuracy
DCA
99.99A ±(2%+0.5A)
999.9A ±(2%+5dgt)*
ACA
0.10–99.99A 50–60Hz±(2%+5dgt)*
>60–400Hz±(2.5%+5
dgt)*
999.9A
VFD
ACA
0.10–99.99A 50–60Hz±(2%+5dgt)*
999.9A
*Formeasuredvalues<1000dgt,add5dgttotheaccuracy.
Overloadprotection:1000A
rms
.
Positionerror:±1%ofreading.
ACconversiontypeandadditionalaccuracyissameasfortheACvoltage.
AC+DCA
rms
accuracy:SameasACAspecification+DCAspecification.TheDCAisaffectedby
the
temperatureandtheresidualmagnetism.UsetheDCAzerofunctiontocompensate.
Table7.3Peakhold:peakmaximum/peakminimum(AConly,NonTRMS)
Function Range Accuracy
ACV
140.0V
±(3.0%+15dgt)
1400V
ACA
140.0A
±(3.5%+15dgt)
1400A
Overloadprotection:1000V
rms
,1000A
rms
Accuracydefinedforsinewaves,ACV>5V
rms
/ACA≥5A
rms
,frequency50–400Hz.Suitablefor
repetitiveevents.