EasyManua.ls Logo

FLIR CM85 - Page 63

FLIR CM85
70 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FLIRCM82USERMANUALDocumentIdentifier:CM82enUS_AD
27
Table7.2Current(TRMS)
Function Range Accuracy
DCA
99.99A ±(2%+0.5A)
599.9A ±(2%+5dgt)*
ACA
0.10–99.99A 50–60Hz±(2%+5dgt)*
>60–400Hz±(2.5%+5
dgt)*
599.9A
VFD
ACA
0.10–99.99A 50–60Hz±(2%+5dgt)*
599.9A
*Formeasuredvalues<1000dgt,add5dgttotheaccuracy.
Overloadprotection:600A
rms
.
Positionerror:±1%ofreading.
ACconversiontypeandadditionalaccuracyissameasfortheACvoltage.
AC+DCA
rms
accuracy:SameasACAspecification+DCAspecification.TheDCAisaffectedbythe
temperatureandtheresidualmagnetism.UsetheDCAzerofunctiontocompensate.
Table7.3Peakhold:peakmaximum/peakminimum(AConly,NonTRMS)
Function Range Accuracy
ACV
140.0V
±(3.0%+15dgt)
1400V
ACA
140.0A
±(3.5%+15dgt)
850A
Overloadprotection:1000V
rms
,600A
rms
Accuracydefinedforsinewaves,ACV>5V
rms
/ACA≥5A
rms
,frequency50–400Hz.Suitablefor
repetitiveevents.

Table of Contents

Related product manuals