EasyManua.ls Logo

Fluke 190-104 - Page 34

Fluke 190-104
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Fluke 190-104/190-204
Service Information
4-16
ADC
B
NORMAL SCOPE
FLUKE 5500A CALIBRATOR
PM9091
CONNECT TO CHANNEL A, B, C, D IN PARALLEL
50 OHM TERMINATIONNO
PM9093
PM9092
Figure 4-5. 5500A Scope Output to Test Tool Input A, B, C, D
2. Select the following Test Tool setup:
Recall the created setup (e.g. SETUP 1, see section 4.4.3): press
SAVE
,
F2
RECALL , select SCREEN+SETUP 1 , press
ENTER
to recall the setup .
Press
SCOPE
Press
F1
READINGS ON.
Press
F2
READING ... and select with
F1
READINGS and with
the arrow keys:
Reading 1, on A, Hz,
Reading 2, on B, Hz,
Reading 3, on C, Hz,
Reading 4, on D, Hz.
3. Press
A
: using
mV
RANGE
and
V
RANGE
select range 50 mV/div for A.
4. Using
TIMEsns
select the required time base setting.
5. Set the 5500A to source a sine wave according to the first test point in Table 4-6.
As no 50Ω termination is applied, the 5500 leveled sine wave output amplitude will
be twice the set value.
6. Observe reading
A and check to see if it is within the range shown under the
appropriate column.
7. Continue through the test points.
8. Next check channel B, C or D in succession:
Press
TRIGGER
and select B, C or D as trigger source with
F1
,
Press B, C or D to assign vertical ‘range’ and ‘move’ to channel B, C or D,
Observe reading B, C or D
9. When you are finished, set the 5500A to 0 (zero) Volt, and to Standby.
Table 4-6. Input A, B, C, D Frequency Measurement Accuracy Test
Model Time
base
5500A-SC... MODE Voltage Frequency Input A&B Reading
all 20 ms/div wavegen, sine 600 mVpp 16 Hz 15.90 to 16.10

Table of Contents

Related product manuals