EasyManua.ls Logo

Fluke 190-104 - Page 49

Fluke 190-104
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Verification
4.6 Scope Input A, B, C, D Tests 4
4-31
1. Connect the Test Tool to the calibrator as shown in Figure 4-19.
ADC
B
NORMAL
SCOPE
FLUKE 5500A CALIBRATOR
PM9091
PM9585 - 50 OHM
USE 50 OHM
TERMINATION
Figure 4-19. Test Tool Input A to TV Signal Generator
2. Select the following Test Tool setup:
Reset the Test Tool
Press
TRIGGER
, then press
F4
to open the Trigger Options menu.
Choose
VIDEO on A... , then from the shown opened menu choose
Polarity: POSITIVE | PAL ( or NTSC or PALplus or SECAM )
Press
F2
to select ALL LINES
Press
F3
to enable the arrow keys for selecting the video line number.
Using
select line number:
622 for PAL, PALplus, or SECAM
525 for NTSC.
Using
mV
RANGE
and
V
RANGE
set the Input A sensitivity to 2 V/div (the actual probe
setting is 10:1).
Using
TIMEsns
select the time base to 20 μs/div.
3. Set the calibrator to mode video with amplitude +100%. Set format and marker line
number to :
PAL 622 (even), for PAL and PALplus
SECAM 622 (even), for SECAM
NTSC 262 even, for NTSC.
4. Observe the trace, and check to see if the Test Tool triggers on the negative pulse
before the marker pulse (see Figure 4-20).
5. Using
select Test Tool line number:
310 for PAL, PALplus or SECAM
262 for NTSC
6. Set the calibrator format and marker line number to :
PAL 310 (odd), for PAL and PALplus

Table of Contents

Related product manuals