Multi-Product Calibrator
Service Information
9
Service Information
Each Model 5500A Calibrator is warranted to the original purchaser for a period of 1
year beginning on the date received. The warranty is located at the front of this manual.
To locate an authorized service center, call Fluke using any of the phone numbers listed
below, or visit us on the World Wide Web: www.fluke.com
USA: 1-888-99-FLUKE (1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-675-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
Or, visit Fluke’s Web site at www.fluke.com.
To register your product, visit register.fluke.com.
After-warranty service is available, but you may choose to repair the calibrator using the
information in the Troubleshooting Chapter of the 5500A Service Manual and the
Module Exchange Program. Refer to the Fluke catalog or contact a Technical Service
Center representative for the module exchange procedure.
Placement and Rack Mounting
You may place the calibrator on a bench top or mount it in a standard-width, 24-inch
(61-cm) deep equipment rack. For bench-top use, the calibrator is equipped with non-
slipping, non-marring feet. To mount the calibrator in an equipment rack, use the 5500A
Rack Mount Kit, Model Y5537. Instructions for rack mounting the calibrator are packed
with the rack mount kit.
Cooling Considerations
Warning
To avoid risk of injury, never operate or power the 5500A
Calibrator without the fan filter in place.
CAUTION
Damage caused by overheating may occur if the area around
the air intake is restricted, the intake air is too warm, or the air
filter becomes clogged.
Baffles direct cooling air from the fan throughout the chassis to internally dissipate heat
during operation. The accuracy and dependability of all internal parts of the calibrator
are enhanced by maintaining the coolest possible internal temperature. You can lengthen
the life of the calibrator and enhance its performance by observing the following rules:
• The area around the air filter must be at least 3 inches from nearby walls or rack
enclosures.
• The exhaust perforations on the sides of the calibrator must be clear of obstructions.
• The air entering the instrument must be at room temperature: make sure the exhaust
air from another instrument is not directed into the fan inlet.
• Clean the air filter every 30 days or more frequently if the calibrator is operated in a
dusty environment. (See the 5500A Operator Manual Chapter 7, “Maintenance” for
instructions on cleaning the air filter.)