Multi-Product Calibrator
Specifications
21
Temperature Calibration (Thermocouple) Specifications
Ran
e
°C
tcal + 5°C
+
°C
1
TC Type Minimum Maximum Source/Measure
1
ear
B 600°C 800°C0.42°C0.44°C
800°C 1000°C0.34°C0.34°C
1000°C 1550°C0.30°C0.30°C
1550°C 1820°C0.26°C0.33°C
C0°C 150°C0.23°C0.30°C
150°C 650°C0.19°C0.26°C
650°C 1000°C0.23°C0.31°C
1000°C 1800°C0.38°C0.50°C
1800°C 2316°C0.63°C0.84°C
E -250°C -100°C0.38°C0.50°C
-100°C-25°C0.12°C0.16°C
-25°C 350°C0.10°C0.14°C
350°C 650°C0.12°C0.16°C
650°C 1000°C0.16°C0.21°C
J -210°C -100°C0.20°C0.27°C
-100°C-30°C0.12°C0.16°C
-30°C 150°C0.10°C0.14°C
150°C 760°C0.13°C0.17°C
760°C 1200°C0.18°C0.23°C
K -200°C -100°C0.25°C0.33°C
-100°C-25°C0.14°C0.18°C
-25°C 120°C0.12°C0.16°C
120°C 1000°C0.19°C0.26°C
1000°C 1372°C0.30°C0.40°C
L -200°C -100°C0.37°C0.37°C
-100°C 800°C0.26°C0.26°C
800°C 900°C0.17°C0.17°C
N -200°C -100°C0.30°C0.40°C
-100°C-25°C0.17°C0.22°C
-25°C 120°C0.15°C0.19°C
120°C 410°C0.14°C0.18°C
410°C 1300°C0.21°C0.27°C
R0°C 250°C0.48°C0.57°C
250°C 400°C0.28°C0.35°C
400°C 1000°C0.26°C0.33°C
1000°C 1767°C0.30°C0.40°C
S0°C 250°C0.47°C0.47°C
250°C 1000°C0.30°C0.36°C
1000°C 1400°C0.28°C0.37°C
1400°C 1767°C0.34°C0.46°C
T -250°C -150°C0.48°C0.63°C
-150°C0°C0.18°C0.24°C
0°C 120°C0.12°C0.16°C
120°C 400°C0.10°C0.14°C
U -200°C0°C0.56°C0.56°C
0°C 600°C0.27°C0.27°C
[1] Does not include thermocouple error.
Notes
1. Resolution is 0.01°C.
2. The 10 µV/°C linear output mode has the same uncertainty as the 300 mV dc range.
3. Applies to both simulated thermocouple output and thermocouple measurement.
4. Temperature standard ITS-90 or IPTS-68 is selectable.