EasyManua.ls Logo

Fluke 5500A - AC Voltage (Sinewave) Specifications

Fluke 5500A
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5500A
Getting Started
16
AC Voltage (Sinewave) Specifications
Ranges Frequency
Absolute Uncertainty, tcal + 5°C
+ (% of output + µV) Resolution
Maximum
Burden
90 days 1 year [1]
1.0 to 32.999 mV 10 to 45 Hz 0.26% 20 µV 0.35% 20 µV1 µV 50
45 Hz to 10 kHz 0.11% 20 µV 0.15% 20 µV
10 to 20 kHz 0.15% 20 µV 0.2% 20 µV
20 to 50 kHz 0.19% 20 µV 0.25% 20 µV
50 to 100 kHz 0.26% 33 µV 0.35% 33 µV
100 to 500 kHz 0.75% 60 µV 1% 60 µV
33 to 329.999 mV 10 to 45 Hz 0.19% 50 µV 0.25% 50 µV1 µV 50
45 Hz to 10 kHz 0.04% 20 µV 0.05% 20 µV
10 to 20 kHz 0.08% 20 µV 0.1% 20 µV
20 to 50 kHz 0.12% 40 µV 0.16% 40 µV
50 to 100 kHz 0.17% 170 µV 0.24% 170 µV
100 to 500 kHz 0.53% 330 µV 0.7% 330 µV
0.33 to 3.29999 V 10 to 45 Hz 0.11% 250 µV 0.15% 250 µV 10 µV 10 mA
45 Hz to 10 kHz 0.02% 60 µV 0.03% 60 µV
10 to 20 kHz 0.06% 60 µV 0.08% 60 µV
20 to 50 kHz 0.10% 300 µV 0.14% 300 µV
50 to 100 kHz 0.17% 1700 µV 0.24% 1700 µV
100 to 500 kHz 0.38% 3300 µV 0.5% 3300 µV
3.3 to 32.9999 V 10 to 45 Hz 0.11% 2500 µV 0.15% 2500 µV 100 µV 10 mA
45 Hz to 10 kHz 0.03% 600 µV 0.04% 600 µV
10 to 20 kHz 0.06% 2600 µV 0.08% 2600 µV
20 to 50 kHz 0.14% 5000 µV 0.19% 5000 µV
50 to 100 kHz 0.17% 17000 µV 0.24% 17000 µV
33 to 329.999 V 45 Hz to 1 kHz 0.04% 6.6 mV 0.05% 6.6 mV 1 mV 5 mA, except
1 to 10 kHz 0.06% 15 mV 0.08% 15 mV 20 mA for
10 to 20 kHz 0.07% 33 mV 0.09% 33 mV 45 to 65 Hz
330 to 1020 V 45 Hz to 1 kHz 0.04% 80 mV 0.05% 80 mV 10 mV 2 mA, except
1 to 5 kHz 0.15% 100 mV 0.20% 100 mV 6 mA for
5 to 10 kHz 0.15% 500 mV 0.20% 500 mV 45 to 65 Hz
5725A Amplifier
100 to 1020 V 45 Hz to 1 kHz 0.04 80 mV 0.05 80 mV 10 mV 50 mA
1 to 20 kHz 0.06 100 mV 0.08 100 mV 10 mV 70 mA
20 to 30 kHz 0.08 100 mV 0.10 100 mV 10 mV 70 mA
100 to 750 V 30 to 100 kHz 0.38 500 mV 0.5 500 mV 10 mV 70 mA
Auxiliary Output [dual output mode only] [2]
10 to 329.999 mV 10 to 20 Hz 0.15 370 µV 0.2 370 µV1 µV5 mA
20 to 45 Hz 0.08 370 µV 0.1 370 µV
45 Hz to 1 kHz 0.08 370 µV 0.1 370 µV
1 to 5 kHz 0.15 450 µV 0.2 450 µV
5 to 10 kHz 0.3 450 µV 0.4 450 µV
0.33 to 3.29999 V 10 to 20 Hz 0.15 450 µV 0.2 450 µV 10 µV5 mA
20 to 45 Hz 0.08 450 µV 0.1 450 µV
45 Hz to 1 kHz 0.07 450 µV 0.09 450 µV
1 to 5 kHz 0.15 1400 µV 0.2 1400 µV
5 to 10 kHz 0.3 1400 µV 0.4 1400 µV
[1] Remote sensing is not provided. Output resistance is < 5 m for outputs > 0.33 V. The AUX output
resistance is < 1 . The maximum load capacitance is 500 pF, subject to the maximum burden current
limits.
[2] There are two channels of voltage output. The maximum frequency of the dual output is 10 kHz.

Table of Contents