26
Power and Dual Output Limit Specifications
Frequency
Voltages
(NORMAL)
Currents
Voltages
(AUX)
Power Factor
(PF)
DC 0 to +1020 V 0 to + 11 A 0 to + 3.3 V
10 to 45 Hz 33 mV to 32.9999 V 3.3 mA to 2.19999 A 10 mV to 3.3 V 0 to 1
45 to 65 Hz 33 mV to 1020 V 3.3 mA to 11 A 10 mV to 3.3 V 0 to 1
65 to 500 Hz 330 mV to 1020 V 33 mA to 2.19999 A 100 mV to 3.3 V 0 to 1
65 to 500 Hz 3.3V to 1020 V 33 mA to 11 A 100 mV to 3.3 V 0 to 1
500 to 1 kHz 330 mV to 1020 V 33 mA to 11 A 100 mV to 3.3 V 1
1 to 5 kHz 3.3V to 1020 V [1] 33 mA to 2.19999 A 100 mV to 3.3 V [1] 1
5 to 10 kHz 3.3V to 1020 V [2] 33 to 329.99 mA 1 to 3.3 V [2] 1
[1] In dual volts, voltage is limited to 3.3 to 500 V in the NORMAL output.
[2] In dual volts, voltage is limited to 3.3 to 250 V in the NORMAL output.
Notes
1. The range of voltages and currents shown in “DC Voltage Specifications,” “DC
Current Specifications,” “AC Voltage (Sinewaves) Specifications,” and “AC Current
(Sinewave) Specifications” are available in the power and dual output modes (except
minimum current for ac power is 0.33 mA). However, only those limits shown in
this table are specified. See “Calculating Power Uncertainty” to determine the
uncertainty at these points.
2. The phase adjustment range for dual ac outputs is 0 to +179.99 degrees.
3. The phase resolution for dual ac outputs is 0.02 degree.