EasyManua.ls Logo

Fluke 5500A - AC Current (Sinewave) Specifications

Fluke 5500A
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5500A
Getting Started
18
AC Current (Sinewave) Specifications
Ranges Frequency
Absolute Uncertainty, tcal + 5°C
+ (% of output + µA) Resolution Compliance
Maximum
Inductive
90 days 1 year Voltage Load
0.029 to 10 to 20 Hz 0.19% 0.15 µA 0.25% 0.15 µA 0.01 µA 3.0 V rms 1 µH
0.32999 mA 20 to 45 Hz 0.09% 0.15 µA 0.125% 0.15 µA
45 Hz to 1 kHz 0.09% 0.25 µA 0.125% 0.25 µA
1 to 5 kHz 0.30% 0.15 µA 0.4% 0.15 µA
5 to 10 kHz 0.94% 0.15 µA 1.25% 0.15 µA
0.33 to 10 to 20 Hz 0.15% 0.3 µA 0.2% 0.3 µA 0.01 µA 3.0 V rms 1 µH
3.2999 mA 20 to 45 Hz 0.08% 0.3 µA 0.1% 0.3 µA
45 Hz to 1 kHz 0.08% 0.3 µA 0.1% 0.3 µA
1 to 5 kHz 0.15% 0.3 µA 0.2% 0.3 µA
5 to 10 kHz 0.45% 0.3 µA 0.6% 0.3 µA
3.3 to 10 to 20 Hz 0.15% 3 µA0.2% 3 µA0.1 µA 3.0 V rms 200 µH,
32.999 mA 20 to 45 Hz 0.08% 3 µA0.1% 3 µA 10 to 500 Hz
45 Hz to 1 kHz 0.07% 3 µA 0.09% 3 µA
1 to 5 kHz 0.15% 3 µA0.2% 3 µA1 µH,
5 to 10 kHz 0.45% 3 µA0.6% 3 µA 500 Hz to 10 kHz
33 to 10 to 20 Hz 0.15% 30 µA 0.2% 30 µA1 µA 3.0 to 2.0 V rms 200 µH,
329.99 mA 20 to 45 Hz 0.08% 30 µA 0.1% 30 µA [1] 10 to 500 Hz
45 Hz to 1 kHz 0.07% 30 µA 0.09% 30 µA
1 to 5 kHz 0.15% 30 µA 0.2% 30 µA5 µH,
5 to 10 kHz 0.45% 30 µA 0.6% 30 µA 500 Hz to 10 kHz
0.33 to
10 to 45 Hz
0.15% 300 µA 0.2% 300 µA 10 µA 3.0 to 2.0 V rms 200 µH,
2.19999 A
45 Hz to 1 kHz
0.08% 300 µA 0.1% 300 µA[2]
45 to 500 Hz
1 to 5 kHz 0.7% 300 µA 0.75% 300 µA5 µH,
500 Hz to 5 kHz
2.2 to 11 A 45 to 65 Hz 0.05% 2000 µA 0.06% 2000 µA 100 µA 2.8 to 1.25 V rms 200 µH,
65 to 500 Hz 0.08% 2000 µA 0.10% 2000 µA [3] 45 to 65 Hz
500 Hz to 1 kHz 0.25% 2000 µA 0.33% 2000 µA1 µH,
65 Hz to 1 kHz
5725A Amplifier
1.5 to 11 A 45 to 1 kHz 0.08% 100 µA 0.1% 100 µA 100 µA 3.0 V rms 400 µH
1 to 5 kHz 0.19% 5000 µA 0.25% 5000 µA
5 to 10 kHz 0.75% 10000 µA 1% 10000 µA
[1] The actual voltage compliance (V
c
) is a function of current output (l
o
), and is given by the formula
V
c
= -3.37*l
o
+3.11. The highest compliance voltage is limited to 3.0 V.
[2] The actual voltage compliance (V
c
) is a function of current output (l
o
), and is given by the formula
V
c
= -0.535*l
o
+3.18. The highest compliance voltage is limited to 3.0 V.
[3] The actual voltage compliance (V
c
) is a function of current output (l
o
), and is given by the formula
V
c
= -0.176*l
o
+3.19. The highest compliance voltage is limited to 2.8 V.
Note
1. 0.30 mA - 2.19999 A can be sent from the 5500A Calibrator to the 5725A
Amplifier front panel terminals. 0.30 mA - 32.999 mA is simultaneously available at the
5500A AUX terminals when sent from the 5500A Calibrator to the 5725A Amplifier
front panel terminals
.