Multi-Product Calibrator
Specifications
19
AC Current (Sinewave) Specifications (cont)
Ranges Frequency
Maximum Distortion and Noise
10 Hz to 100 kHz
Bandwidth
+ (% output + µA)
0.02 to 0.32999 mA 10 to 20 Hz 0.15% + 1.0 µA
20 to 45 Hz 0.1% + 1.0 µA
45 Hz to 1 kHz 0.05% + 1.0 µA
1 to 5 kHz 0.5% + 1.0 µA
5 to 10 kHz 1.0% + 1.0 µA
0.33 to 3.2999 mA 10 to 20 Hz 0.15% + 1.5 µA
20 to 45 Hz 0.06% + 1.5 µA
45 Hz to 1 kHz 0.02% + 1.5 µA
1 to 5 kHz 0.5% + 1.5 µA
5 to 10 kHz 1.2% + 1.5 µA
3.3 to 32.999 mA 10 to 20 Hz 0.15% + 5 µA
20 to 45 Hz 0.05% + 5 µA
45 Hz to 1 kHz 0.07% + 5 µA
1 to 5 kHz 0.3% + 5 µA
5 to 10 kHz 0.7% + 5 µA
33 to 329.99 mA 10 to 20 Hz 0.15% + 50 µA
20 to 45 Hz 0.05% + 50 µA
45 Hz to 1 kHz 0.07% + 50 µA
1 to 5 kHz 0.2% + 50 µA
5 to 10 kHz 0.7% + 50 µA
0.33 to 2.19999 A
10 to 45 Hz 0.2% + 500 µA
45 Hz to 1 kHz 0.1% + 500 µA
1 to 5 kHz 1.4% + 500 µA
2.2 to 11 A 45 to 65 Hz 0.2% + 5 mA
65 to 500 Hz 0.1% + 5 mA
500 Hz to 1 kHz 0.4% + 5 mA
5725A Amplifier
1.5 to 11 A 45 Hz to 1 kHz 0.05% + 1 m A
1 to 5 kHz 0.12% + 1 mA
5 to 10 kHz 0.5% + 1 mA