Multi-Product Calibrator
Specifications
15
Resistance Specifications
Ranges
Absolute Uncertainty, tcal + 5°C
+ (% of output + ) [2]
Resolution
Allowable Current
[1]
90 days 1 year [4]
0 to 10.99 0.009% 0.008 [3] 0.012% 0.008 [3] 0.001 1 to 125 mA
11 to 32.999 0.009% 0.015 [3] 0.012% 0.015 [3] 0.001 1 to 125 mA
33 to 109.999 0.007% 0.015 [3] 0.009% 0.015 [3] 0.001 1 to 70 mA
110 to 329.999 0.007% 0.015 [3] 0.009% 0.015 [3] 0.001 1 to 40 mA
330 to 1.09999 k 0.007% 0.06 0.009% 0.06 0.01 250 µA to 18 mA
1.1 to 3.29999 k 0.007% 0.06 0.009% 0.06 0.01 250 µA to 5 mA
3.3 to 10.9999 k 0.007% 0.6 0.009% 0.6 0.1 25 µA to 1.8 mA
11 to 32.9999 k 0.007% 0.6 0.009% 0.6 0.1 25 µA to 0.5 mA
33 to 109.999 k 0.008% 6 0.011% 6 1 2.5 µA to 0.18 mA
110 to 329.999 k 0.009% 6 0.012% 6 1 2.5 µA to 0.05 mA
330 k to 1.09999 M 0.011% 55 0.015% 55 10 250 nA to 0.018 mA
1.1 to 3.29999 M 0.011% 55 0.015% 55 10 250 nA to 5 µA
3.3 to 10.9999 M 0.045% 550 0.06% 550 100 25 nA to 1.8 µA
11 to 32.9999 M 0.075% 550 0.1% 550 100 25 nA to 0.5 µA
33 to 109.999 M 0.4% 5.5 k 0.5% 5.5 k 1000 2.5 nA to 0.18 µA
110 to 330 M 0.4% 16.5 k 0.5% 16.5 k 1000 2.5 nA to 0.06 µA
[1] Continuously variable from 0 to 330 M.
[2] Applies for COMP OFF (to the 5500A Calibrator front panel NORMAL terminals) and 2-wire and 4-wire
compensation.
[3] The floor adder is improved to 0.006 (0 to 10.99 range) and 0.010 (11 to 329.999 ) if the 5500A
Calibrator is zeroed (ohms zero or instrument zero) within 8 hours and temperature is +1°C of zeroing
ambient temperature.
[4] Do not exceed the largest current for each range. For currents lower than shown, the floor adder increases
by Floor
(new)
= Floor
(old)
x I
min
/I
actual
. For example, a 100 µA stimulus measuring 100 has a floor uncertainty
of 0.01 x 1 mA/100 µA = 0.1 .
Ranges Maximum Voltage [1] Maximum Lead Resistance [2]
0 to 10.99 1.37 V <3.2
11 to 32.999 4.1 V <3.2
33 to 109.999 7.7 V <3.2
110 to 329.999 13.2 V <3.2
330 to 1.09999 k 19.8 V <6
1.1 to 3.29999 k 16.5 V <6
3.3 to 10.9999 k 19.8 V <6
11 to 32.9999 k 16.5 V <6
33 to 109.999 k 19.8 V <6
110 to 329.999 k 16.5 V (n/a 110 k and above)
330 k to 1.09999 M 19.8 V (n/a 110 k and above)
1.1 to 3.29999 M 16.5 V (n/a 110 k and above)
3.3 to 10.9999 M 19.8 V (n/a 110 k and above)
11 to 32.9999 M 16.5 V (n/a 110 k and above)
33 to 109.999 M 19.8 V (n/a 110 k and above)
110 to 330 M 19.8 V (n/a 110 k and above)
[1] This is for the largest resistance for each range. The maximum voltage for other values is Imax (highest
value of Allowable Current above) multiplied by Rout.
[2] Maximum lead resistance for no additional error in 2-wire COMP.