5500A
Getting Started
20
Capacitance Specifications
Ranges
Absolute Uncertainty, tcal + 5°C
+ (% of output + ηF) Resolution Allowed
Typical
Frequency
90 days 1 year
Frequency
for <1%
Error
0.33 to 0.4999 ηF 0.38% 0.01 ηF 0.5% 0.01 ηF 0.1 pF 50 to 1000 Hz 10 kHz
0.5 to 1.0999 ηF 0.38% 0.01 ηF 0.5% 0.01 ηF 0.1 pF 50 to 1000 Hz 10 kHz
1.1 to 3.2999 ηF 0.38% 0.01 ηF 0.5% 0.01 ηF 0.1 pF 50 to 1000 Hz 10 kHz
3.3 to 10.999 ηF 0.38% 0.01 ηF 0.5% 0.01 ηF 1 pF 50 to 1000 Hz 10 kHz
11 to 32.999 ηF 0.19% 0.1 ηF 0.25% 0.1 ηF 1 pF 50 to 1000 Hz 10 kHz
33 to 109.99 ηF 0.19% 0.1 ηF 0. 25% 0.1 ηF 10 pF 50 to 1000 Hz 10 kHz
110 to 329.99 ηF 0.19% 0.3 ηF 0.25% 0.3 ηF 10 pF 50 to 1000 Hz 10 kHz
0.33 to 1.0999 µF 0.19% 1 ηF 0.25% 1 ηF 100 pF 50 to 1000 Hz 5 kHz
1.1 to 3.2999 µF 0.26% 3 ηF 0.35% 3 ηF 100 pF 50 to 1000 Hz 2 kHz
3.3 to 10.999 µF 0.26% 10 ηF 0.35% 10 ηF1 ηF 50 to 400 Hz 1.5 kHz
11 to 32.999 µF 0.30% 30 ηF 0.40% 30 ηF1 ηF 50 to 400 Hz 800 Hz
33 to 109.99 µF 0.38% 100 ηF 0.50% 100 ηF 10 ηF 50 to 200 Hz 400 Hz
110 to 329.99 µF 0.50% 300 ηF 0.70% 300 ηF 10 ηF 50 to 100 Hz 200 Hz
330 to 1.1 mF 1% 300 ηF 1% 300 ηF 100 ηF 50 to 100 Hz 150 Hz
Notes
1. Specifications apply to both dc charge/discharge capacitance meters and ac RCL
meters.
2. The output is continuously variable from 330 pF to 1.1 mF.
3. For all ranges, the maximum charge and discharge current is 150 mA peak or 30 mA
rms. The peak voltage is 4 V, except the 330 µF - 1.1 mF range is limited to 1 V.
The maximum lead resistance for no additional error in 2-wire COMP mode is 10
ohms.