Fluke 5500A - AC Power (45 Hz to 65 Hz) Specification Summary, PF=1

Fluke 5500A
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5500A
Getting Started
24
DC Power Specification Summary
5500A Calibrator Current Range
Voltage Range 3.3 to 8.999 mA 9 to 32.999 mA 33 to 89.99 mA 90 to 329.99 mA
Absolute Uncertainty, tcal + 5°C, + (% of watts output)
90 days 33 mV to 1020 V 0.03% 0.02% 0.03% 0.02%
1 year 33 mV to 1020 V 0.04% 0.03% 0.04% 0.03%
5500A Calibrator Current Range
Voltage Range 0.33 to 0.8999 A 0.9 to 2.1999 A 2.2 to 4.4999 A 4.5 to 11 A
Absolute Uncertainty, tcal + 5°C, + (% of watts output)
90 days 33 mV to 1020 V 0.07% 0.05% 0.08% 0.06%
1 year 33 mV to 1020 V 0.08% 0.06% 0.12% 0.09%
5725A Amplifier Current Range
Voltage Range 1.5 to 4.4999 A 4.5 to 11 A
Absolute Uncertainty, tcal + 5°C, + (% of watts output)
90 days 33 mV to 1020 V 0.09% 0.07%
1 year 33 mV to 1020 V 0.10% 0.08%
Note
1. To determine dc power uncertainty with more precision, see the individual “DC
Voltage Specifications” and “DC Current Specifications” and “Calculating Power
Uncertainty.”