EasyManua.ls Logo

Fluke 5500A - AC Voltage, Triangle Wave Characteristics (Typical); AC Current (Sinewave) Extended Bandwidth Specifications; AC Current (Non-Sinewave) Specifications

Fluke 5500A
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5500A
Getting Started
34
AC Voltage, Triangle wave Characteristics (typical)
Linearity to 1 kHz Aberrations
0.3% of p-p value, from 10% to 90% point <1% of pk-to-pk value, with amplitude >50% of range
AC Current (Sinewave) Extended Bandwidth Specifications
Ranges Frequency
1-Year Absolute Uncertainty,
tcal + 5°C,
+ (% of output + % of range) [2]
Maximum
Current
Resolution
%Output %Range
All current ranges, <330 mA 0.01 - 10 Hz 5.0% 0.5% 2 digits each
range
10 - 10 kHz See “AC Current (Sinewave) Specifications”
AC Current (Non-Sinewave) Specifications
Trianglewave &
Truncated Sinewave
Frequency
1-Year Absolute Uncertainty,
tcal + 5°C,
+ (% of output + % of range) [2]
Maximum
Current
Ranges [1]
%Output %Range
Resolution
2.9 to 92.999 mA 0.01 to 10 Hz 5.0% 0.5% Two digits, e.g., 75 mA
10 to 45 Hz 0.25% 0.5%
45 to 1 kHz 0.25% 0.25% Six digits on each
range
1 to 10 kHz 5.0% 0.5%
93 to 929.999 mA 0.01 to 10 Hz 5.0% 0.5% Two digits
10 to 45 Hz 0.25% 0.5%
45 to 1 kHz 0.25% 0.5% Six digits on each
range
1 to 10kHz 5.0% 1.0%
0.93 to 6.19999 A 10 to 45 Hz 5.0% 1.0% Two digits
45 to 1kHz 0.5% 0.5% Six digits on each
range
1 to 5 kHz 5.0% 1.0%
6.2 to 31 A 45 to 500 Hz 2.0% 0.5% Two digits on each
range
500 to 1 kHz 5.0% 1.0% Six digits on each
range
[1] All waveforms are peak-to-peak output ranges.
[2] Uncertainty is stated in peak-to-peak. Amplitude is verified using an rms-responding DMM.

Table of Contents

Other manuals for Fluke 5500A

Related product manuals