5500A
Getting Started
12
General Specifications
Warmup Time Twice the time since last warmed up, to a maximum of 30 minutes
Settling Time 5 seconds typical for all functions and ranges
Internal Diagnostics Comprehensive analog and digital self-tests on demand. Self-tests are
valid after warm-up time.
Standard Interfaces IEEE-488 (GPIB), RS-232, 5725A Amplifier
Temperature Performance Operating: 0°C to 50°C
Calibration (tcal): 15°C to 35°C
Storage: -20°C to 70°C
Electromagnetic
Compatibility
Designed to operate in Standard Laboratory environments where the
Electromagnetic environment is highly controlled. If used in areas with
Electromagnetic fields of 1 to 3 V/m, resistance outputs have a floor
adder of 0.508 . Performance not specified above 3 V/m.
Temperature Coefficient Temperature Coefficient for temperatures outside tcal +5°C is 0.1X/°C of
the 90-day specification (or 1-year, as applicable).
Relative Humidity Operating: <80% to 30°C, <70% to 40°C,<40% to 50°C
Storage: <95%, noncondensing
Altitude Operating: 3,050 m (10,000 ft) maximum
Nonoperating: 12,200 m (40,000 ft) maximum
Safety Designed to comply with IEC 1010-1 (1992-1); ANSI/ISA-S82.01-1994;
CAN/CSA-C22.2 No. 1010.1-92
Analog Low Isolation 20V
EMI/RFI Designed to comply with FCC Rules Part 15; VFG 243/1991
Line Power Line Voltage (selectable): 100 V, 120 V, 220 V, 240 V
Line Frequency: 47 to 63 Hz
Line Voltage Variation: +10% about line voltage setting
Power Consumption 5500A Calibrator, 300 VA; 5725A Amplifier, 750 VA
Dimensions
5500A Calibrator:
Height, 17.8 cm (7 inches), standard rack increment, plus 1.5 cm
(0.6 inch) for feet on bottom of unit;
Width, 43.2 cm (17 inches), standard rack width
Depth, 47.3 cm (18.6 inches) overall.
5725A Amplifier:
Height, 13.3 cm (5.25 inches)
Width, 43.2 cm (17 inches), standard rack width
Depth, 63.0 cm (24.8 inches) overall.
Weight 5500A Calibrator:
20 kg (44 pounds)
5725A Amplifier:
32 kg (70 pounds)