AC Voltage (Non-Sinewave) Specifications
Trianglewave &
Truncated Sine
Ranges
Frequency
1-Year Absolute Uncertainty,
tcal + 5°C,
+ (% of output + % of range) [2]
Maximum
Voltage
Resolution
Peak-to-Peak [1]
%Output %Range
Normal Channel (Single Output Mode)
2.9 to 92.999 mV 0.01 to 10 Hz 5.0% 0.5% Two digits on each range
93 to 929.999 mV 10 to 45 Hz 0.25% 0.5%
0.93 to 9.29999 V 45 to 1 kHz 0.25% 0.25% Six digits on each range
9.3 to 92.9999 V 1 to 20 kHz 0.5% 0.25%
20 to 100 kHz 5.0% 0.5%
Auxiliary Output (Dual Output Mode)
0.01 to 10 Hz 5.0% 0.5% Two digits on each range
93 to 929.999 mV 10 to 45 Hz 0.25% 0.5%
0.93 to 9.29999 V 45 to 1 kHz 0.25% 0.25% Six digits on each range
1 to 10 kHz 5.0% 0.5%
[1] To convert peak-to-peak to rms for trianglewave, multiply the peak-to-peak value by 0.2886751. To
convert peak-to-peak to rms for truncated sinewave, multiply the peak-to-peak value by 0.2165063.
[2] Uncertainty is stated in peak-to-peak. Amplitude is verified using an rms-responding DMM.
Squarewave
Ranges (p-p) [1]
Frequency
1-Year Absolute Uncertainty,
tcal + 5°C
+ (% of output + % of range) [2]
Maximum Voltage
Resolution
%Output %Range
Normal Channel (Single Output Mode)
2.9 to 65.999 mV 0.01 to 10 Hz 5.0% 0.5% Two digits on each range
66 to 659.999 mV 10 to 45 Hz 0.25% 0.5%
0.66 to 6.59999 V 45 to 1 kHz 0.25% 0.25% Six digits on each range
6.6 to 65.9999 V 1 to 20 kHz 0.5% 0.25%
20 to 100 kHz 5.0% 0.5%
Auxiliary Output (Dual Output Mode)
0.01 to 10 Hz 5.0% 0.5% Two digits on each range
66 to 659.999 mV 10 to 45 Hz 0.25% 0.5%
0.66 to 6.59999 V 45 to 1 kHz 0.25% 0.25% Six digits on each range
1 to 10 kHz 5.0% 0.5%
[1] To convert peak-to-peak to rms for squarewave, multiply the peak-to-peak value by 0.5000000.
[2] Uncertainty is stated in peak-to-peak. Amplitude is verified using an rms-responding DMM.