Multi-Product Calibrator
Specifications
13
DC Voltage Specifications
Absolute Uncertainty, tcal + 5°C Stability
Maximum
Ranges
+ (% of output + µV)
24 hours, + 1°C
Resolution
Burden
90 days 1 year
+ (ppm output + µV)
[1]
0 to 329.9999 mV 0.005% 3 µV 0.006% 3 µV 5 ppm + 1 µV0.1 µV 50
0 to 3.299999 V 0.004% 5 µV 0.005% 5 µV 4 ppm + 3 µV1 µV 10 mA
0 to 32.99999 V 0.004% 50 µV 0.005% 50 µV 4 ppm + 30 µV 10 µV 10 mA
30 to 329.9999 V 0.0045% 500 µV 0.0055% 500 µV 4.5 ppm + 300 µV 100 µV5 mA
100 to 1020.000 V 0.0045% 1500 µV 0.0055% 1500 µV 4.5 ppm + 900 µV 1000 µV5 mA
Auxiliary Output (dual output mode only) [2]
0 to 329.999
mV
0.03% 350
µV
0.04% 350
µV
30 ppm + 100 µV1 µV5 mA
0.33 to 3.3 V 0.03% 350
µV
0.04% 350
µV
30 ppm + 100 µV 10 µV5 mA
[1] Remote sensing is not provided. Output resistance is < 5 m for outputs > 0.33 V. The AUX output has an output
resistance of < 1.
[2] Two channels of dc voltage output are provided.
Noise
Ranges Bandwidth
0.1 to 10 Hz
Bandwidth
10 to 10 kHz
p-p rms
+ (ppm output + µV) µV
0 to 329.9999 mV 1 µV4 µV
0 to 3.299999 V 10 µV 50 µV
0 to 32.99999 V 100 µV 600 µV
30 to 329.9999 V 10 ppm + 1 mV 20 mV
100 to 1020.000 V 10 ppm + 5 mV 20 mV
Auxiliary Output (dual output mode only) [1]
0 to 329.999 mV 5 µV 20 µV
0.33 to 3.3 V 20 µV 200 µV
[1] Two channels of dc voltage output are provided.