Multi-Product Calibrator
Additional Specifications
29
Additional Specifications
The following paragraphs provide additional specifications for the 5500A Calibrator ac
voltage and ac current functions. These specifications are valid after allowing a warm-up
period of 30 minutes, or twice the time the 5500A has been turned off. All extended
range specifications are based on performing the internal zero-cal function at weekly
intervals, or when the ambient temperature changes by more than 5°C. (See Chapter 4,
“Front Panel Operations.”)
Frequency Specifications
Frequency Resolution 1-Year Absolute Uncertainty, Jitter
Range tcal + 5°C
.01 to 119.99 Hz .01 Hz 25 ppm, + 1 mHz 2 µs
120.0 to 1199.9 Hz 0.1 Hz 25 ppm, + 1 mHz 2 µs
1.200 to 11.999 kHz 1.0 Hz 25 ppm, + 1 mHz [1] 2 µs
12.00 to 119.99 kHz 10 Hz 25 ppm, + 15 mHz 140 ns
120.0 to 1199.9 kHz 100 Hz 25 ppm, + 15 mHz 140 ns
1.200 to 2.000 MHz 1 kHz 25 ppm, + 15 mHz 140 ns
[1] ± (25 ppm + 15 mHz) above 10 kHz