Multi-Product Calibrator
Specifications
11
Specifications
The following paragraphs describe the details for the 5500A specifications. All
specifications are valid after allowing a warm-up period of 30 minutes, or twice the time
the 5500A has been turned off. (For example, if the 5500A has been turned off for 5
minutes, the warm-up period is 10 minutes.)
All specifications apply for the temperature and time period indicated. For temperatures
outside of tcal + 5°C (tcal is the ambient temperature when the 5500A was calibrated),
the temperature coefficient is less than 0.1 times the 90-day specifications per °C
(limited to 0°C - 50°C). These specifications also assume the 5500A Calibrator is zeroed
every seven days or when the ambient temperature changes more than 5°C. (See
“Zeroing the Calibrator” in Chapter 4 of the Operator Manual.)
Also see additional specifications later in this chapter for information on extended
specifications for ac voltage and current. The dimensional outline for the 5500A
Calibrator is shown in Figure 6.
POWER
I
O
0•
123
456
789
ENTER
M
k
m
VHz
FIELD
EDIT
/
+
F
OPR EARTH SCOPE BOOST MENU
PREV
SHIFT
RESET
CE
SETUP
REF
NEW
TC
MEAS
¡F
µ
n
p
W
dBm sec
¡CA
MULT
x
DIV
÷
OUT
TRIG
5500A
CALIBRATOR
20V PK
MAX
HI
LO
TC
TRIG
OUT
1000V
RMS
MAX
20V
RMS
MAX
1V PK
MAX
20V PK
MAX
NORMAL AUX
SCOPE
V, ,
RTD
A, -SENSE,
AUX V
200V PK
MAX
STBY
43.2 cm (17 in)
47.0 cm (18.5 in) 6.4 cm
(2.5 in)
For Cable
Access
17.8 cm
(7 in)
F1-04.eps
Figure 6. 5500A Calibrator Dimensional Outline