Multi-Product Calibrator
Additional Specifications
33
AC Voltage, DC Offset Specifications
Range [1]
(Normal Channel) Offset Range [2]
Maximum
Peak
1-Year Absolute Offset
Uncertainty, tcal + 5°C [3]
Signal
+ (% Output (dc) + µV)
Sinewaves
3.3m to 32.999 mV 0 to 50 mV 80 mV 0.1% + 33 µV
33m to 329.999 mV 0 to 500 mV 800 mV 0.1% + 330
0.33 to 3.29999 V 0 to 5 V 8 V 0.1% + 3300
3.3 to 32.9999 V 0 to 50 V 55 V 0.1% + 33 mV
Trianglewaves and Truncated Sinewaves
9.3 to 92.999 mV p-p 0 to 50 mV 80 mV 0.1% + 93 µV
93 to 929.999 mV p -p 0 to 500 mV 800 mV 0.1% + 930
0.93 to 9.29999 V p-p 0 to 5 V 8 V 0.1% + 9300
9.3 to 92.9999 V p-p 0 to 50 V 55 V 0.1% + 93 mV
Squarewaves
6.6 to 65.999 mV p-p 0 to 50 mV 80 mV 0.1% + 66 µV
66 to 659.999 mV p-p 0 to 500 mV 800 mV 0.1% + 660
0.66 to 6.59999 V p-p 0 to 5 V 8 V 0.1% + 6600
6.6 to 65.9999 V p-p 0 to 50 V 55 V 0.1% + 66 mV
[1] Offsets are not allowed on ranges above the highest range shown above.
[2] The maximum offset value is determined by the difference between the peak value of the selected
voltage output and the allowable maximum peak signal. For example, a 10 V peak-to-peak square
wave output has a peak value of 5 V, allowing a maximum offset up to +50 V to not exceed the 55 V
maximum peak signal. The maximum offset values shown above are for the minimum outputs in each
range.
[3] For frequencies 0.01 to 10 Hz, and 500 kHz to 2 MHz, the offset uncertainty is 5% of output, + 1% of
the offset range.
AC Voltage, Squarewave Characteristics
Risetime
@ 1 kHz
Typical
Settling
Time
@ 1 kHz
Typical
Overshoot
@ 1 kHz
Typical
Duty Cycle Range Duty Cycle Uncertainty [1]
<1 µs <10 µs to 1%
of final value
<2% 1% to 99%, <3.3 V p-p,
0.01 Hz to 100 kHz
+ (0.8% of period +140 ns)
for frequencies >10 kHz;
+ (0.8% of period + 2 µs)
for frequencies <10 kHz.
[1] For duty cycles of 10.00% to 90.00%.