EasyManua.ls Logo

Fluke 5500A - DC Power Specification Summary

Fluke 5500A
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Multi-Product Calibrator
Specifications
23
Temperature Calibration (RTD) Specifications (cont.)
RTD Type
Range (°C) Absolute Uncertainty, tcal + 5°C, +
(°C) [1]
Minimum Maximum 90 days 1 year
Pt 385, 1000 -200°C-80°C0.03°C0.03°C
-80°C0°C0.03°C0.03°C
0°C 100°C0.03°C0.04°C
100°C 260°C0.04°C0.05°C
260°C 300°C0.05°C0.06°C
300°C 400°C0.05°C0.07°C
400°C 600°C0.06°C0.07°C
600°C 630°C0.22°C0.23°C
PtNi 385, 120 -80°C0°C0.06°C0.08°C
(Ni120) 0°C 100°C0.07°C0.08°C
100°C 260°C0.13°C0.14°C
Cu 427, 10 [2] -100°C 260°C0.3°C0.3°C
[1] Applies for COMP OFF (to the 5500A Calibrator front panel NORMAL terminals) and 2-wire and 4-
wire compensation.
[2] Based on MINCO Application Aid No. 18.
Note
1. Resolution is 0.003°C.

Table of Contents

Other manuals for Fluke 5500A