EasyManua.ls Logo

Fluke 5500A - AC Current (Sinewave) Specifications

Fluke 5500A
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Multi-Product Calibrator
Specifications
17
AC Voltage (Sinewave) Specifications (cont)
Ranges Frequency
Maximum Distortion and Noise
10 Hz to 5 MHz Bandwidth
+ (% output + µV)
1.0 to 32.999 mV 10 to 45 Hz 0.15% + 90 µV
45 Hz to 10 kHz 0.035% + 90 µV
10 to 20 kHz 0.06% + 90 µV
20 to 50 kHz 0.15% + 90 µV
50 to 100 kHz 0.25% + 90 µV
100 to 500 kHz 0.3% + 90 µV
33 to 329.999 mV 10 to 45 Hz 0.15% + 90 µV
45 Hz to 10 kHz 0.035% + 90 µV
10 to 20 kHz 0.06% + 90 µV
20 to 50 kHz 0.15% + 90 µV
50 to 100 kHz 0.20% + 90 µV
100 to 500 kHz 0.20% + 90 µV
0.33 to 3.29999 V 10 to 45 Hz 0.15% + 200 µV
45 Hz to 10 kHz 0.035% + 200 µV
10 to 20 kHz 0.06% + 200 µV
20 to 50 kHz 0.15% + 200 µV
50 to 100 kHz 0.20% + 200 µV
100 to 500 kHz 0.20% + 200 µV
3.3 to 32.9999 V 10 to 45 Hz 0.15% + 2 mV
45 Hz to 10 kHz 0.035% + 2 mV
10 to 20 kHz 0.08% + 2 mV
20 to 50 kHz 0.2% + 2 mV
50 to 100 kHz 0.5% + 2 mV
33 to 329.999 V 45 Hz to 1 kHz 0.15% + 10 mV
1 to 10 kHz 0.05% + 10 mV
10 to 20 kHz 0.6% + 10 mV
330 to 1020 V 45 Hz to 1 kHz 0.15% + 30 mV
1 to 10 kHz 0.07% + 30 mV
5725A Amplifier
100 to 1020 V 45 Hz to 1 kHz 0.07%
1 to 20 kHz 0.15%
20 to 30 kHz 0.3%
100 to 750 V 30 to 100 kHz 0.4%
Auxiliary Output [dual output mode only] 10 Hz to 100 kHz Bandwidth
10 to 329.999 mV 10 to 20 Hz 0.2% + 200 µV
20 to 45 Hz 0.06% + 200 µV
45 Hz to 1 kHz 0.08% + 200 µV
1 to 5 kHz 0.3% + 200 µV
5 to 10 kHz 0.6% + 200 µV
0.33 to 3.29999 V 10 to 20 Hz 0.2% + 200 µV
20 to 45 Hz 0.06% + 200 µV
45 Hz to 1 kHz 0.08% + 200 µV
1 to 5 kHz 0.3% + 200 µV
5 to 10 kHz 0.6% + 200 µV