26
Press FORMAT NV MEM in the utility functions
menu to open the following menu:
oe11f.eps
ALL replaces the contents of the EEPROM with
factory defaults. The rear panel CALIBRATION
switch must be in the ENABLE position.
CAL replaces all calibration constants with factory
defaults. The rear panel CALIBRATION switch
must be in the ENABLE position.
SETUP replaces the setup parameters with factory
defaults (see the Factory Defaults table on the
next page).