59
Capacitance Output
To set synthesized capacitance output at
Calibrator NORMAL terminals:
1. Press R to clear the Calibrator output.
2. Connect to the UUT.
Note
Since this is a synthesized output, the
terminal connections from the Calibrator to
the UUT must be LO to LO and HI to HI.
3. Set the UUT to measure capacitance.
4. Enter the desired capacitance output (up to 5
digits).
5. Press a multiplier key. For example, b then
cfor µF.
6. Press F.
7. Control Display shows entered value. For
example, 123.45 µF (below).
oe50f.eps
8. Press E.
oe51f.eps