53
Dual AC Voltage Output
Note
Tie the NORMAL LO and AUX LO
terminals together at the UUT or at the
Calibrator, via the “LO”s tied softkey.
Selecting voltage in units of dBm is not allowed in
dual ac voltage mode. To set dual ac voltages at
Calibrator NORMAL and AUX terminals:
1. Press R to clear the Calibrator output.
2. Connect to the UUT.
3. Set the UUT to measure dual ac voltage.
4. Enter the desired voltage output at NORMAL
terminals (up to 6 digits).
5. Press a multiplier key, if necessary. For
example, press c.
6. Press V.
Note
The AUX output is limited to 5 V rms for
sinewaves, 14 V p-p for squarewaves,
triangle and truncated sinewaves.
7. Enter the desired voltage output at AUX
terminals (up to 6 digits).
8. Press a multiplier key, if necessary.
9. Press V.
10. Enter the desired frequency output (up to 5
digits).