Setting a Phase Angle
To set a phase shift in degrees for dual ac voltage
or ac power output:
1. Press WAVE MENUS to open waveform menu.
2. Press PHASE to open phase entry menu.
3. Enter the desired phase angle (up to 5 digits).
4. Press I to select leading (+) or lagging ()
phase shift.
5. Control Display shows entries. For example, a
leading phase angle of 123.45 degrees. Press
E.
oe68f.eps
6. Press Pone or more times to return to
previous menus.
Setting a Power Factor
To set a phase shift as a power factor (PF), where
PF=Cosine Φ, and Φ is phase shift:
1. Press WAVE MENUS to open waveform menu.
2. Press PHASE to open phase entry menu.
3. Press SHOW PF to open power factor entry
menu.
4. Enter the desired power factor (3 digits).