Resistance Output
To set synthesized resistance output at Calibrator
NORMAL terminals:
1. Press R to clear the Calibrator output.
2. Connect to the UUT.
Note
Terminal connections from Calibrator to
UUT must be LO to LO and HI to HI.
3. Set the UUT to measure resistance.
4. Enter the desired resistance (up to 6 digits).
5. Press a multiplier key, if necessary.
6. Press Q.
7. Control Display shows entered value. For
example, 12.3456 k.
oe47f.eps
8. Press E.
oe48f.eps