60
9. Press O to activate the Calibrator output.
oe52f.eps
COMP (Compensation) applies two-wire
compensation or turns compensation off for
capacitance of 110 nF and above. This feature
cancels out lead resistance, NOT capacitance.
Temperature Simulation (TC)
Note
Make sure the thermocouple wire and
plug are not affected by extraneous
temperature sources. For example, do not
place your fingers on the thermocouple
plug or wire when simulating a
thermocouple.
To set a simulated thermocouple temperature
output at Calibrator front panel TC connector:
1. Press R to clear the Calibrator output.
2. Connect to the UUT.
Note
Use thermocouple wire and miniconnectors
that match the type of thermocouple.
3. Set the UUT to measure temperature.
4. Enter the desired temperature output (up to 6
digits).
5. Press the C key for °C, or b C for °F.