61
6. The Control Display shows the entered value.
For example, 123.456 °C.
oe53f.eps
7. Press E.
oe54f.eps
8. Press O to activate the Calibrator output.
Note
The temperature is cleared to 0
°
C (32
°
F) if
you change between tc and rtd, or change
the type of thermocouple (except a B-type
thermocouple, which starts at 600
°
C).