63
Out@TC terminal displays dc voltage at TC
terminals.
TC MENUS opens submenus for thermocouple
outputs.
OUTPUT selects the temperature device:
thermocouple (tc) or resistance temperature
detector (rtd).
TYPE selects the thermocouple type to simulate.
UNITS selects °C or °F temperature unit.
REF SRC selects intrnl (Internal) or extrnl
(External) reference source.
REF shows the value of the temperature
reference.
TYPE selects the thermocouple type to simulate.
Temperature Simulation (RTD)
To set a simulated RTD temperature output at
Calibrator NORMAL terminals:
1. Press R to clear the Calibrator output.
2. Connect to the UUT.
3. Set the UUT to measure temperature.
4. Enter the desired temperature output (up to 6
digits).
5. Press the C key for °C, or b Cfor °F.
6. Control Display shows entered value. For
example, 123.456 °C.
oe53f.eps

Table of Contents

Other manuals for Fluke 5520A