64
7. Press E.
oe54f.eps
8. Press O to activate the Calibrator output.
Note
The temperature is cleared to 0
°
C (32
°
F)
if you change between tc and rtd, or
change the type of rtd.
oe58f.eps
Output @ NORMALdisplays location of output
terminals (always NORMAL).
TYPE selects rtd curve.
OUTPUT selects temperature device:
thermocouple (tc) or resistance temperature
detector (rtd).
COMP applies four-wire compensation, two-wire
compensation or turns compensation off. For a 3-
wire RTD configuration, turn compensation off.