Setting Outputs
DC Voltage Output
To set a dc voltage output at the Calibrator
NORMAL terminals:
1. Press R to clear the Calibrator output.
2. Connect to the UUT.
3. Set the UUT to measure dc voltage.
4. Enter the desired voltage output (up to 7
digits).
5. Press I to select polarity.
6. Press a multiplier key, if necessary.
7. Press V.
8. The Control Display shows entered value. For
example, 123.4567 mV.
oe26f.eps
9. Press E. The output value appears on the
Output Display.
oe27f.eps
10. Press O to activate the Calibrator output.
11. Press the auto/locked softkey to toggle
between autorange mode or lock the Calibrator
in the present range.