Fluke 5520A - 3-14. Capacitance Calibration

Fluke 5520A
88 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
78
5. Press PF to toggle a leading (lead) or lagging
(lag) power factor. For example, a leading
power factor of .678. Press E.
oe69f.eps
6. Press Pone or more times to return to
previous menus.
Setting a DC Offset
To set a dc offset for single ac voltage outputs:
oe70f.eps
1. Press WAVE to select desired waveform.
2. Press OFFSET to open offset entry menu.
3. Enter the desired offset. For example, 0.123 V.
Press E.
oe71f.eps