40
AC Voltage Output
To set ac voltage output in volts or as a power
output in dBm (referenced to a selectable load
impedance) at the Calibrator NORMAL terminals:
1. Press R to clear the Calibrator output.
2. Connect to the UUT.
3. Set the UUT to measure ac voltage.
4. Enter the desired voltage output (up to 6
digits).
5. Press a multiplier key, if necessary.
6. Press V(volts) or b V (dBm).
If you are using dBm, a softkey appears that lets
you choose an impedance value from a list.
7. Control Display shows entered value. For
example, 2.44949 V.
oe28f.eps
8. Enter the desired frequency output (up to 5
digits).
9. Press a multiplier key, if necessary.
10. Press H..
11. The Control Display shows entered value. For
example, 1.1234 kHz (below).
oe29f.eps