45
DC Power Output
Note
Tie the NORMAL LO and AUX LO
terminals together at the UUT or at the
Calibrator, via the “LO”s “tied” softkey.
To set dc a voltage at Calibrator NORMAL
terminals and a dc current on AUX or 20A
terminals:
1. Press R to clear the Calibrator output.
2. Connect to the UUT.
3. Set the UUT to measure dc power.
4. Enter the desired voltage output (up to 7
digits).
5. Press I to select polarity.
6. Press a multiplier key, if necessary.
7. Press V .
8. Enter the desired current output (up to 6 digits).
9. Press I to select polarity.
10. Press a multiplier key, if necessary.
11. Press A.