AC Current Output
To set an ac current output at Calibrator AUX or
20A terminals:
1. Press R to clear the Calibrator output.
2. Connect to the UUT.
3. Set the UUT to measure ac current.
4. Enter the desired current output (up to 6 digits).
5. Press a multiplier key, if necessary.
6. Press A.
7. Enter the desired frequency (up to 5 digits).
8. Press a multiplier key, if necessary.
9. Press H.
10. The Control Display shows the entered value.
For example, 123.456 mA and 1.1234 kHz.
oe34f.eps
11. Press E.
oe35f.eps
10. Press O to activate the Calibrator output.